E-mail
info@chinyee.cn
-
Phone
15810002992
-
Address
1505, Building A, Caizhi International Building, No. 18 Zhongguancun East Road, Haidian District, Beijing
Beijing Chengyi Hengyi Technology Co., Ltd
info@chinyee.cn
15810002992
1505, Building A, Caizhi International Building, No. 18 Zhongguancun East Road, Haidian District, Beijing
German MI VM-3000 mercury analyzerIntroduction:
VM-3000 mercury analyzerIt can continuously monitor the mercury concentration in the gas,It can be used in the laboratory or installed anywhere where mercury monitoring is required. Industrial practitioners, researchers from research institutes, or safety officers can all operate proficiently. This instrument has powerful functions and wide applications, such as workplace monitoring, exhaust gas emission monitoring, chemical emission monitoring, ground air screening in polluted areas, quality control of hydrogen and other gases, laboratory application testing, etc.
German MI VM-3000 mercury analyzerApplication field:
Occupational Health and Health Geological Environment Monitoring Workplace Monitoring Mercury Leakage Screening Investigation of Polluted Sites Environmental Monitoring Hazardous Waste Inspection
Measurement principle of VM-3000
The sample gas is introduced into an optical sample cell through a maintenance free diaphragm pump. The UV beam irradiates the sample cell, and some of the ultraviolet light is absorbed by the mercury atoms in the sample. This method is called "Atomic Absorption Spectroscopy", abbreviated as AAS. This method has high selectivity and sensitivity. Compared with other early methods of mercury measurement, AAS method has its own advantages and importance. AAS method has low interference, does not require the step of adding mercury amalgam, and does not require expensive carrier gas.
VM-3000 Analysis Performance
VM-3000 uses a high-frequency driven infinitely low pressure mercury lamp as the ultraviolet light source. This type of lamp emits light with an extremely narrow bandwidth, which is in phase with the absorption spectrum of Hg atoms, thereby minimizing the factors that affect sensitivity. The UV light source in VM-3000 adopts a reference beam feedback control method, therefore it has * stability. The total background noise is below 0.1 µ g/m ³.
Design Features of VM-3000
Metal casing with sturdy handle
Optional accessories: Bracket version with 19 "mounting bracket
Diaphragm pumps have a long service life
It is easy to replace the filter membrane of the injection port filter
Both optical and electronic systems have undergone factory calibration and have stable performance, without the need for frequent calibration. (Provide short-term factory recalibration services)
Technical parameters of VM-3000
Measurement principle: |
UV absorption |
Wavelength: |
253,7nm |
UV light source: |
Ultra low pressure mercury lamp (EDL) |
Measurement unit: |
Fused silica, heated to avoid condensation |
Stability method: |
Control the light source through reference beam feedback; Automatic zeroing |
Measuring range |
Range I: 0.1-100 µ g/m ³ (0.01-10 ppb) Scope II: 0-1000 µ g/m ³ Range III: 0-2000 µ g/m ³ (0-200 ppb) |
Sensitivity: |
0.1µg/m³(0.01 ppb) |
response time (99% signal): |
Continuous operation and real-time reading of values |
Alarm method: |
Level 3 programmable |
Status alarm: |
Measurement unit contaminated, battery status (option: UV light source exhausted) |
Display mode |
Graphic LCD display with background lighting |
Signal output: |
RS-232 interface for connecting to PC or printer, data recording function (optional) |
Battery options: |
External 12V DC power supply; Optional 12V integrated battery, with a working time of 5h |