HI97701 Italian HANNA residual chlorine analyzer


Measurement range: 0.00 to 5.00 mg/L (ppm) CI2
Reminder: In measurement mode, press 【 Methods 】 to enter the reagent type selection. Press the 【 ▼▲ 】 key to choose between powder reagent 【 Free Chlorine (Powder) 】 or water-based reagent 【 Free Chlorine (Liquid) 】 to meet your needs for different measurement unit results.
Meets IP67 waterproof standard, adopts * * optical system design, with * * accuracy, repeatability, and measurement time required, CAL Check ™ The function allows for performance validation and calibration of instruments using NIST traceable standards, with multiple measurement methods available. Users can choose to use powder or water reagents, and a built-in reaction timer ensures consistency between sample measurements and user results. It automatically records 50 measurement readings, displays the date and time of one user calibration according to Good Laboratory Practice (GLP), has a battery status indicator light, and an automatic shutdown setting
HI97701 Italian HANNA residual chlorine analyzerTechnical Specifications:
technical indicators |
HI97701 residual chlorine |
|
Measurement Mode |
measurement range |
0.00 to 5.00 mg/L(ppm)CI2 |
resolution |
0.01 mg/L(≤3.50ppm)、0.10 mg/L(>3.50ppm) |
|
精度@25 ºC/77ºF |
Reading ± 3% ± 0.03 mg/L (ppm) |
|
standard method |
USEPA method 330.5, DPD method |
|
optical system |
Light source mode |
silicon photocell, LED @ 525 nm |
Filter bandwidth |
8 nm ±1.0 nm |
|
Other indicators |
data management |
Automatically store, query and extract 50 measurement readings |
Power mode |
3 × 1.5AAA high-performance battery, automatically shuts down after 15 minutes of inactivity in measurement mode |
|
Usage environment |
0 to 50 ° C (32 to 122 ° F), RH max *, compliant with IP67 waterproof standard |
|
Dimensions & Weight |
Host size: 142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0 '), host weight: 380 g (13.4 oz.) |
|
Exclusive attachments |
Glass colorimetric dish |
HI731397 glass colorimetric dish (HI731331+HI731336), specification: outer diameter 24.6mm (inner diameter 22mm), 4 sets/group |
Measurement reagents |
Residual chlorine reagent (powder): HI93701-01 residual chlorine DPD reagent, prediction times: 100 times; HI93701-03 residual chlorine DPD reagent, predicted times: 300 times |
|
Residual chlorine reagent (water agent): HI93701-F residual chlorine DPD reagent, prediction times: 300 times |
||
Standard Curve Verification Group |
HI97701-11 residual chlorine NIST measurement curve calibration group, standard values @ 25 º C: 0.00 and 1.00 ± 0.03mg/L (ppm) CI2 |
|