HI97740 Hanna Nickel Ion Portable Detector


measurement range:0.000 to 1.000 mg/L (as Ni)
Kind Reminder: Sample * Temperature Range:20 to 24ºC,When the temperature of the test sample is higher than30ºCWhen there is a possibility of turbidity, it may occur2-3dropHI93703-51Customized special dispersant [optional] added separately to the glass colorimetric dish, gently rotated until turbidity is eliminated.
Compliant with IP67 waterproof standards, featuring innovative optical system design with excellent accuracy, repeatability, and measurement time, Hanna's carefully designed CAL Check ™ The function allows for performance validation and calibration of instruments using NIST traceable standards, with multiple measurement methods available. Users can choose to use powder or water reagents, and a built-in reaction timer ensures consistency between sample measurements and user results. It automatically records 50 measurement readings, displays the date and time of one user calibration according to Good Laboratory Practice (GLP), has a battery status indicator light, and an automatic shutdown setting.
HI97740 microcomputer nickel low range waterproof photometric analyzer |
HI96740Microcomputer nickel low range photometer |
measurement range:0.000 to 1.000 mg/L (as Ni) |
measurement range:0.000 to 1.000 mg/L (as Ni) |
Optical System: Silicon Photovoltaic Cell, LED @ 575 nm, Filter bandwidth: 8 nm ± 1.0 nm |
Optical system: Narrowband interference filter, silicon photovoltaic cellTungsten lamp @ 575 nm |
GLP function: Automatically record, query, and clear * calibration value and date/time once |
GLP function: Manually record, query, and clear * one calibration value and date/time |
Data storage: automatic storage, query extraction * 50Secondary measurement reading |
data storage: No such function available |
display mode128 x 64 pixel black and white LCD display with backlight |
Display mode:none |
HI97740 Hanna Nickel Ion Portable DetectorTechnical Specifications:
technical indicators |
HI97740nickel |
|
Measurement Mode |
measurement range |
0.000 to 1.000 mg/L (as Ni) |
resolution |
0.001 mg/L |
|
精度@25 ºC/77ºF |
Reading±7% ± 0.010 mg/L |
|
standard method |
Refer to the 1- (2-pyridylazo) -2-naphthol PAN method |
|
optical system |
Light source mode |
silicon photocell,LED @ 575 nm |
Filter bandwidth |
8 nm ±1.0 nm |
|
Other indicators |
data management |
50Reading storage |
Power mode |
3 × 1.5AAA high-performance battery, in measurement mode, 15Automatically shut down after minutes of inactivity |
|
Usage environment |
0 to 50°C(32 to 122°F),RH max *,Compliant with IP67Waterproof standard |
|
Dimensions & Weight |
Host size: 142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0 '), host weight: 380 g (13.4 oz.) |
|
Exclusive attachments |
Glass colorimetric dish |
HI731397 glass colorimetric dish(HI731331+HI731336)Specification: Outer diameter 24.6mm (inner diameter 22mm), 4Set/Group |
Measurement reagents |
HI93740-01 nickel low range reagent, prediction times: 50 times; HI93740-03 Nickel Low Range Reagent, Prediction Times: 150time |
|
Standard Curve Verification Group |
HI97740-11 Nickel NIST Measurement Curve Calibration Group, Standard Value @ 25 º C:0.00 and 0.500 ± 0.020 mg / L(ppm)Ni/ |
|