Sievers Eclipse Bacterial endotoxin testing (BET) platform
Compliant, clear, and consistent bacterial endotoxin testing equipment and services
The Eclipse endotoxin detection system can achieve automation and high accuracy in endotoxin detection without affecting compliance, biochemical reactions, or increasing laboratory footprint. By using embedded endotoxin standards and positive product controls (PPC) with less than 30 pipetting steps, the Sievers Eclipse microplate obtained can achieve automated processes without the complexity or cost of robotic endotoxin testing equipment. In addition to automated endotoxin testing, Sievers analyzers also provide customer service to ensure the success of laboratory testing.
Only 9 minutes are needed to achieve compliant testing settings for 21 samples
Reduce the usage of horseshoe crab reagent by 90%
Meets all requirements of the pharmacopoeia: USP<85>, EP 2.6.14, JP 4.01, and ChP
Embedded endotoxin standard curve and positive product control (PPC)
Provide segmented standard curves, samples, and positive product controls (PPC)
Sensitivity as low as 0.005 EU/mL
Control the reaction temperature at 37+/-1 C using a cultivation absorbance analyzer
Reduce the need for extensive operator training and the risk of repetitive stress injuries
Increase daily sample completion volume
Use FDA approved commercially available horseshoe crab reagents provided by major horseshoe crab reagent manufacturers
Enterprise software complies with 21 CFR PART 11 and data reliability guidelines
System Specifications | |
|---|---|
|
Bacterial endotoxin detection method |
Dynamic color rendering |
|
Detection Mode |
absorbance |
|
Range, ppm |
0.005–50 EU/mL |
|
accuracy |
≤ 15% CV onset time |
|
accuracy |
50-200% of the true value |
|
Sample Type |
Water solution, injected with a pipette |
|
calibration |
Up to 12 months |
|
analysis time |
Up to 2 hours |
|
Sample temperature |
37 ± 1 °C |
|
ambient temperature |
17–30 °C |
|
capacity |
Standard configuration: single sample flow; Optional: Dual Sample Flow |
|
temperature control |
37 ± 0.5 °C |
|
light source |
LED emitter |
|
Fluid fault detection |
1450 nm emitter |
|
Optical accuracy |
≤ 5% deviation from expected value |
|
Optical linearity |
R ≥ 0.980 |
|
Optical wavelength filter |
405 nm |
|
read interval |
5 seconds |
Analyzer specifications | |
|
output |
USB data output |
|
display |
OLED |
|
Power requirements |
100–240 Volts AC @50/60 Hz |
|
fuse |
T8 A 250 VAC fuse, size 5 x20 millimeters. Littlfuse218008 or Cooper Bussmann S 506-8-R. |
|
size |
Height: 17.5 cm (6.9 in), Width: 35.1 cm (13.8 in), Depth: 50.3 cm (19.8 in) |
|
weight |
10 kg (22.1 lbs) |
|
Safety Certification |
UL 61010-1:2012Ed.3+R:20Apr2016 CSA C22.2#61010-2-020 CSA C22.2#61010-1-12:2012Ed.3+U1; U2 IEC 61010-1:2010 Ed.3+C1; C2 UL 61010-2-010:2015Ed.3 IEC 61010-2-010:2003Ed.2 UL 61010-2-020:2016Ed.3 IEC 61010-2-020:2016Ed.3 CSA C22.2#61010-2-010:2015Ed.3 |
|
relative humidity |
85%, no condensation |
|
altitude |
3,000 m(9,800 ft) |
|
pollution level |
2 |
Sievers Eclipse Bacterial endotoxin testing (BET) platform