The Sievers M9 TOC analyzer imported by Veolia provides stable and accurate data for monitoring, cleaning validation, and other applications of ultrapure water and pharmaceutical water. It uses membrane conductivity technology to enhance its analytical performance, reliability, and ease of use. Features such as TOC and Level 1 conductivity synchronous testing for pharmaceutical customers, as well as Turbo acceleration mode, make M9 stand out from other similar instruments in the market while meeting regulatory requirements in the pharmaceutical industry, including USP 643 and 645, and EP 2.2.44 Total Organic Carbon.

Product Details

Sievers M9 TOC analyzer imported from Veolia

Sievers M9 Laboratory TOC AnalyzerCharacteristics and advantages of

  • Two minute sample analysis time

  • Detect samples extracted discretely, or can be connected to Sievers automatic sampler for large-scale sample analysis

  • Automatically perform calibration, validation, and data analysis operations

  • The reagent adaptive function can automatically determine the optimal flow rate for each sample

  • Very suitable for various sample matrices and concentrations

  • Very little preventive maintenance (usually only a few hours per year)

  • 12-month calibration stability period

  • Optional "Stage 1" conductivity testing capability for synchronous detection of TOC and conductivity

  • Optional Turbo acceleration mode, analysis time only takes 4 seconds

    Sievers M9 TOC analyzer imported from Veolia

    Operating specifications*

    scope 0.03 ppb to 50 ppm
    Precision and accuracy < 1% RSD
    accuracy ± 2% or ± 0.5 ppb (whichever is higher)
    Display reading 3 significant digits
    Sample Type Automatic sampler or discrete sample aspiration
    calibration The usual stable period is 12 months
    analysis time 2 minutes (4 seconds when using the optional Turbo mode)
    Sample temperature 5-95 °C(41-203 °F)
    ambient temperature 5-40 °C(41-104 °F)
    Instrument samples
    flow velocity
    0.5 mL/min

    Analyzer specifications

    output USB device port (1); USB host port (3); Modbus TCP/IP
    display 7 'WVGA 800x480 pixel, color LCD touch screen
    electricity 100 - 240 V~,50 - 60 Hz,100 VA
    fuse Replace the same type and size of fuse: T 1.6 A 250 VAC fuse (slow melting), size 5 X 20 mm electrical socket
    size H: 42.2 cm(16.6 in),W: 24.6 cm(9.7 in),D: 40.0 cm(15.8 in)
    weight 9.4 kg(20.6 lb)
    Safety Certification ETL, CE (see conformity declaration)

    Sample conductivity (optional)

    Initial conductivity range 0.01 to 2000 μ S/cm
    Conductivity accuracy ± 0.005 μ S/cm or ± 1% (whichever is higher)
    Precision and accuracy <0.25% rsd=''>

    environment

    High relative humidity 0-95%, no condensation
    High altitude 3,000 m(9.800 ft)
    pollution level 2
    The UV lamp in this product contains mercury and must be recycled or disposed of in accordance with local, state, and federal laws.

    *The analytical performance can be achieved under controlled laboratory conditions, minimizing operational and standard errors to the greatest extent possible.

M9 Laboratory Type Options and Accessories

  • Sievers automatic sampler: 63 or 120 sample positions capacity

  • Data Pro 2 software: custom sampling protocol, automatic calibration and verification protocol

  • DataGuard softwareProvide compliance support for 21 CFR Part 11 and Appendix 11

  • Inorganic carbon remover (ICR):Improve the accuracy of TOC in samples with high inorganic carbon content

  • Sample conductivityDetect and report the conductivity of the sample

  • Turbo acceleration mode:Provide a 4-second analysis time