Semiconductor green lightLaser power meter tallprecisionLaser power meterModel: ELD-102
1、 Overview
The ELD-102 series is a new type oftallA sensitive, fast response, low-temperature drift, digital direct reading laser power measuring instrument, equipped with detectors suitable for measuring ranges of 40mW, 200mW, 600mW, and resolutions of 0.01mW and 1mW, respectively. The detector can not only ensure that the testing instrument hastallIt has high sensitivity and can withstand high levels of stresstallThe power density. This series of instruments has been widely used for measuring various laser powers from ultraviolet to ultraviolet infrared.
The ELD-102 series not only has a real-time power measurement method, but also adds two aging curve measurement methods: 1. 10 minute power aging curve 2. 24-hour power aging curve
These two measurement methods are very suitable for assessing the stability of laser output power, and the measurement curve is directly displayed on the local dot matrix LCD screen.
2、 Technical indicators
1. Measurement object: Continuous laser power
2、 响应波长: 445nm, 532nm, 635nm, 660nm, 780nm, 808nm, 850nm, 980nm
3. Sensitive area: Φ 10mm
4. Measure the continuous power meter, and the allowable power density for the detector is 600mW/cm2
5. Range: 0-40 mW, 0-200 mW, 0-600 mW
6. Response time:<0.1s>7. Uncertainty: ± 3%
8. The spectrum is within the range of 445nm to 980nm, and the power can be directly read.
9. Power supply: DC5V
10. Display: 240 × 128 dot matrix LCD screen
11. Power consumption:<1w>Usage conditions: General optical laboratory, avoid strong light exposure.
usage method
1. Power supply
After connecting the power cord to the power socket on the back panel of the instrument, connect the plug to a 220V socket, and then press the power switch. The LED screen will light up, and the measurement can begin.
2. Regarding light
Align the sensitive area of the detection receiver with the measured light source.
3. Real time power measurement mode
After power on, it enters real-time power measurement mode. The lower part of the LCD display indicates the current measurement result, and the upper part indicates the calibration wavelength of this device