Welcome Customer !

Membership

Help

Shanghai Huifen Scientific Analysis Instrument Co., Ltd
Custom manufacturer

Main Products:

ag-mach>Products
Product Categories

Shanghai Huifen Scientific Analysis Instrument Co., Ltd

  • E-mail

    2954198810@qq.com

  • Phone

    13651667376

  • Address

    Room 3661, No. 13, Lane 65, Huandong 1st Road, Fengjing Town, Jinshan District, Shanghai

Contact Now

Content of domestic ethylene oxide standard solution

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

EO standard/ethylene oxide standard solution 10.0mg/mL is used to detect the residual amount of ethylene oxide after one-time ethylene oxide detoxification with a top air chromatograph to determine whether it meets the standard requirements.

Product Details

EO standard/ethylene oxide standard solution 10.0mg/mLSpecification: 10mg/ml

1. Ethylene oxide detection standard/EO standardDilution method:

(1). Take 5ml of pure water and place it in a headspace bottle (take 6 bottles)

(2). Take 0.5ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain 1ppm of the standard sample

(3). Take 1.0ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain a 2ppm standard sample

(4). Take 2.0ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain 4ppm of the standard sample

(5). Take 3.0ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain 6ppm of the standard sample

(6). Take 4.0ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain 8ppm of the standard sample

(7). Take 5.0ul of the standard sample with a 10ul injection needle and add it to a 5ml headspace vial to obtain 10ppm of the standard sample


Draw the standard curve for ethylene oxide:

Dissolve the diluted standard solutions of 6 series of concentrationsQuickly extract the upper gas from the equilibrated standard sample using a headspace sampler and inject it into the gas chromatograph,When the standard sample reaches gas-liquid equilibrium, different concentrations of liquid correspond to different concentrations of gas. Take the equilibrium gas, inject it into the sampler, and record the peak height (or area) of EO. Draw a peak height (or area) - concentration standard curve based on the peak height (or area) corresponding to different concentrations of standard solutions. Take an appropriate amount of sample and prepare it into a sample. Measure the peak height (or area) using the same method, and use the measured peak height (or area) to determine the concentration of the corresponding standard solution from the standard curve. Calculate the residual amount of EO according to the formula.

image:


Ethylene oxide standard solution

Specification: C (C2H4O): 10.0mg/mL

Execution standard: JJF1343

Storage method: 4-10 ℃ refrigerator storage and use

Net content: 10mL/tube