Welcome Customer !

Membership

Help

Ikeda House Industrial (Shenzhen) Co., Ltd
Custom manufacturer

Main Products:

ag-mach>Products
Product Categories

Ikeda House Industrial (Shenzhen) Co., Ltd

  • E-mail

    zx@ikedaya.cn

  • Phone

    17503077078

  • Address

    1207 Zhanyuan Business Building, No. 10 Huishi Street Lane 1, Songhe Community, Longhua Street, Longhua District, Shenzhen

Contact Now

Zirconia Oxygen Concentration Meter Company

NegotiableUpdate on 01/17
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

TORAY Toray Compact Zirconia Oxygen Concentrator RF-400-01 Ikedaya Industrial Advantage Supply, and provide quotation pictures

Product Details

TORAY Compact Zirconia Oxygen Concentration Meter RF-400-01 from Japan

TORAY Compact Zirconia Oxygen Concentration Meter RF-400-01 from Japan


The zirconia oxygen concentration meter RF-400-01 is a product developed for nitrogen reflow soldering equipment, utilizing our years of accumulated manufacturing technology for oxygen concentration meters in the electronics industry. It is also compatible with high concentrations (dirty gases).
Equipped with our compact zirconia sensor, it can perform stable and high-precision measurements for a long time, and is also equipped with an oxygen sensor protection mechanism, achieving durability beyond traditional products.

This is a good oxygen concentration meter that combines simple operation and compact design, and can be integrated into various devices including nitrogen reflux equipment.


Product Features

Adopting a new type of particle filter

It uses a new type of particle filter that can react with oil and color it, making it easy to determine its expected lifespan. The filter can be replaced without removing the oxygen meter from the instrument.

Using special filters

It improves the durability of the oxygen sensor and can handle various deteriorated substances.

Wide power supply

Compatible with AC100V to 240V power supply.

Convenient measurement range

The measurement range is 0-10000ppm, commonly used in nitrogen reflow soldering equipment (0-1% range can also be selected).

Compatible with standards

Safety: CE, UKCA...EN61010-1
EMC:CE…EN61326-1

Environmental regulations

RoHS指令... EN63000

Optional support

The filters we provide help prevent sensor degradation and ensure accurate measurement of indicated and measured values.

① Activated carbon filter: a filter that adsorbs trace amounts of organic matter and small amounts of combustible gases. In addition to preventing impurities from entering the sensor, it can also prevent the measured oxygen from burning inside the sensor, thereby achieving accurate readings.

② Cotton filter: a filter that prevents dust and solid substances from entering. Prevent foreign objects from entering the oxygen concentration meter or adhering to the sensor to prevent deterioration of the oxygen concentration meter.

③ Silicon filter: a filter that adsorbs gases containing silicon. By preventing silicon containing gases from entering the sensor, sensor degradation can be prevented and accurate measurements can be achieved.

Product Specifications

Specifications

shape Desktop/panel mounted type
display mode 4-digit display (concentration display)
measurement range 10ppm~100vol%O2
0- 1/10/100/1000/10000ppm
(1)/10/
measurement range 10 -20 ~10 -0 atm
*Cannot guarantee 0-10ppm range and below
*You can choose between a range of 0-10000ppm or a range of 0-1%
Sampling method Built in pump continuous suction type
gas supply volume 1000-2000mL/min
Sensor gas quantity 60 ± 10 milliliters per minute
Gas connection port INLET (sample gas inlet) Rc1/4
BYPASSOUT (bypass outlet) Rc1/4
OUTLET (sample gas outlet) Rc1/4
Comparing gases atmosphere
a large amount of About 6.5 kilograms
paint color Munsell N3 matte paint (black)

feature

repetitiveness When 1% (10000ppm) ≤ measured gas ≤ 100%: ± 1% FS
When 10ppm ≤ measured gas ≤ 1% (10000ppm): ± 2% FS
(Cannot guarantee display range below 0-10ppm)
Air point stability ± 1% FS/24 hours

engineering specifications

Measurement value output External output: 4-20mA, RS-232C
Concentration alarm output LED display (upper and lower limits)
Contact output: 2a output separately
Contact capacity: AC30V/DC30V, 1A
Self diagnostic function Preheating abnormality, thermocouple abnormality, internal temperature abnormality
CPU abnormality, furnace temperature abnormality
Asymmetric potential abnormality, sensor resistance value abnormality, AIR point calibration abnormality
SPAN point calibration abnormality
Sample gas conditions Must be free of flammable components, halogens, silica, corrosive components, and water droplets
Pressure: 29.4 kPa or lower
Pressure flow rate: 1000 to 2000 mL/min
Temperature: 0 to 50 ° C
Humidity: Dew point lower than ambient temperature
warm-up time Within 6 minutes
power supply Voltage: AC100V to 240V
Required power capacity: 200VA
Power consumption: 50W
Installation requirements Installation location: indoor, non explosion proof area
Environmental temperature: 0 to 50 ° C
Environmental humidity: 20 to 85% RH (non condensing)
Does not contain flammable components, halogens, silica or corrosive components
option Activated carbon filter
Cotton filter cartridge
Silicon filter cartridge