Online ammonia nitrogen automatic monitoring instrumentfeature
The new photoelectric quantitative technology is not affected by interference such as high chromaticity, suspended solids, and bubbles, with high reliability and accuracy, and strong anti-interference ability.
High detection accuracy, low measurement limit, and small long-term drift of the instrument.
It has one click self-test function and self diagnostic function.
Adopting a single channel highly integrated valve group, maintenance and cleaning are simple and fast.
24/7 networking function, monitoring instrument operation status anytime, anywhere.
It has alarm function, quality control function, and reverse control function.
It has automatic cleaning and calibration functions.
Capable of whole point measurement, interval measurement, and external control measurement.
Automatically work after power failure and recovery.
Automatic prompt and reset function for faults and missing reagents (samples).
L has fault recording function.
L can save historical data for more than 3 years.
Ensure separate discharge of cleaning solution and waste liquid
Online ammonia nitrogen automatic monitoring instrument


Technical Specifications
Measurement method: HJ536-2009 water quality ammonia nitrogen salicylic acid spectrophotometric method (customizable Nessler method)
Measurement range: 0-2 mg/L; 0-10 mg/L ; 0-50 mg/L ; Customizable range
Indication error: ≤± 3% Detection limit: 0.01mg/L
Repeatability: ≤± 2% Waste liquid volume: 6mL/time
Memory effect: ≤± 3% Actual water sample comparison: ± 8%
Constant temperature time: 360 seconds Voltage stability: ≤± 5%
Measurement cycle: 30 minutes Calibration cycle: any time
Maintenance cycle: ≥ 720h/time Maintenance workload:<2 hours/month
Reagent consumption: 3 months/500ml Display output: Equipped with a 10.1-inch color LCD touch screen
Color rendering temperature: The recommended temperature is 45 ℃, which can be set according to the actual water sample situation
Data export: Measurement values can be imported into a USB drive through a USB port and saved
Signal output: RS485/RS232/USB interface/standard 2-channel 4-20mA output/standard 2-channel switch input/output
Environmental requirements: Temperature controlled indoor environment, recommended temperature (5-28 ℃), humidity ≤ 90% (no condensation)
Power supply and output: (220V ± 22) V/AC, (50 ± 0.5Hz), 5A, 150W
Instrument size: Upper cabinet 600 * 450 * 300mm; Upper cabinet 700 * 450 * 300mm