Quotation for UV aging test chamber

NegotiableUpdate on 01/18
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This experimental equipment simulates the damage caused by solar ultraviolet light, and conducts urban weather resistance testing on the tested material by exposing it to controlled high temperature light. It uses ultraviolet lamps to simulate the radiation effect of sunlight, and simulates dew and rain through condensation and water spraying. In just a few days or weeks, the ultraviolet irradiation equipment can withstand the damage that takes months or even years to occur outdoors, including fading, color changes, loss of luster, powdering, cracking, wrinkling, foaming, brittleness, strength reduction, oxidation, etc. Its test results can be used to select new materials, improve existing materials, or evaluate changes in material formulations.

Product Details

1Product Name

UV aging test chamber (box type)

IIProduct Model

ST-XS-UV340

IIIEquipment purpose

This experimental equipment simulates the damage caused by solar ultraviolet light by exposing the tested material to controlled high temperature illumination

Conduct urban weather resistance testing on the materials. It uses ultraviolet lamps to simulate the radiation effect of sunlight, and simulates dew through condensation and water spraying

Water and rainwater, only a few days or weeks, UVlineThe irradiation equipment can be used, but now it takes several months outdoors

Damage that takes several years to occur, including fading, color changes, loss of luster, powdering, cracking, and cracking

The test results of wrinkling, foaming, brittleness, strength reduction, oxidation, etc. can be used to select new materials and improve existing materials,

Or evaluate changes in material formulation.

4、 Characteristics of the test chamber

★ Provide comprehensive and detailed real-time monitoring and calibration of important experimental parameters such as irradiance, temperature, and timelineRecord (controller with)

The recorder function can also generate reports and music through a computerlinePrint.

★ It can conduct irradiation/condensation/water spray tests separately, or program any combination of irradiation/condensation/water spray tests.

★ Adopting a self-developed programmable 7-inch true color touch screen controller, it has good stability, monitoring, and operation

Sexuality and maintainability.

★ Imported ultraviolet radiation with expensive materialslineThe lamp provides a lifespan reference (the lamp is a consumable with a lifespan of approximately)1600 hours), square

Users can confirm the replacement time of the lamp tube to save usage costs. Through the irradiance measurement and control device (solar eye), the irradiance can be measured

Measure and control the irradiance to automatically stabilize at the set value, greatly extending the service life of the lamp.

★ Adopting automated water supply and drainage design, combined with high cost high-power mercury water, to ensure the flow rate and uniformity of water spray.

Most of the main components are from international brands, with a small portion from domestic brands, which improves the safety and reliability of the product.

The installation and wiring of the electrical part of the equipment shall be carried out in accordance with international electrical standards.

lThis experimental equipment prohibits:

lTesting and storage of samples of flammable, explosive, and volatile substances

lTesting and storage of corrosive substance samples

lTesting or storage of biological samples

lTesting and Storage of Strong Electromagnetic Emission Source Samples

five. meet the standard

lGB/T16422.3-2022Plastic laboratory light source exposure test methodPart 3: Fluorescent UV Lamp

lGB/T14522-2008 Test Method for Artificial Climate Aging of Plastics, Coatings, and Rubber Materials for Mechanical Industry Products - Fluorescence Ultraviolet

The Lamp "a、 Fluorescent ultraviolet/condensation test method.

lGB/T16585-1996 National Standard of the People's Republic of China - Test for Artificial Climate Aging of Vulcanized Rubber (Fluorescent UV Lamp)

Method ".

lCompliant with international testing standards:Current UV standards such as ISO 4892-3, ISO 11507, SAE J2020, etclineAging test standards.

Note:Other types of standards are not listed in detail. If necessary, please confirm with our technical department.

sixcontrol instrument

adoptself-developedprogrammable7-inch true color touch screen controllerHas good stability, monitoring, and operability

And maintainability.

1. Fixed value experiment: The set values of temperature and humidity can be reached according to a certain time or proportion.

2. Program experiment: This mode allows users to edit specific programs themselves and adjust temperature and humidity according to the program.

3. Operation settings: This function allows users to perform operation related settings.

4. Fault record: This function allows staff to view DI fault records.

5. Appointment setting: This function can facilitate staff to start fixed value tasks or program tasks at regular intervals.

sevenDimensions and Weight

7.1. Inner box size(About)

1000×350×550mm (W* H *D)

7.2Outer box size(About)

1300×1650×650mm (W* H *D)Based on actual dimensions

7.3Weight

about150

eighty percentPerformance indicators

81. TemperaturewetRange of degrees

Temperature range: normal temperature+10~80℃; Blackboard measurement range: 45-80 ℃; Condensation temperature range: 40~60 ℃; Light temperature:

50~70℃; Humidity range:≥90%RH

8.2.temperaturevolatility

volatility±3℃; Resolution: 0.1 ℃

8.3The center distance of the lamp

70mm (center distance between test sample and lamp tube: Plate base frame test plate (adjustable up and down)

8.4.Lamp brand and model

ImportUVA-340 (315-400nm), UVB-313 (280-315nm) (optional), our default standard configuration

6 light tubes. If you need 313 light tubes, please specify before placing an order.

8.5.irradiance

0.35~1.2W/m ² adjustable UVA-340 (315~400nm)

8.6.Temperature control method

PID self-tuning temperature control method

nineIntroduction to functional areas

9.1.Irradiation system(Simulating sunlight)

light source6The rated power isA 40W imported ultraviolet fluorescent lamp with UVA-340 and UVB-313 light sources for users to choose from and configure. like

Like other light tubes,The aging of UV lamps can lead to a decrease in lamp output, and there is no sun eye control system to rotate the position of UV use

The method to compensate. This system can be used for many application standards, but it has inherent limitations. This method of rotating positions

The law cannot compensate for the impact caused by different batches of lamps or different surrounding environments.

9.2.Solar Eye System

Changes in light intensity can affect the rate of material aging, while differences in spectra may simultaneously affect both the rate and type of material aging

Create an impact. Accurate and reproducible test results can only be obtained by controlling irradiance through the solar eye system.

The solar eye system is a precise closed-loop control system that can automatically maintain the set light intensity through a feedback loop. controller

It can continuously monitor the intensity of ultraviolet light and compensate for any aging or other factors caused by the lamp by adjusting the output of the lamp

Changes in light intensity. The reproducibility and repeatability of the test results of the solar eye irradiance control system are better than those of manual irradiance control, and

Can effectively extend the lifespan of the lamp tube. The sun eye is installed in the distribution cabinetUVloopIn the middle, there is a group of two solar eye monitoring devices

The light intensity of the lamp tube is measured and the data is transmitted to the controller. The controller compares the measured irradiance value with the set value and then controls

The device commands the power system to adjust the output of the lamp tube to maintain the set light intensity value.

9.3.UVA-340fluorescent tube

Can emit from315~400nmThe wavelength of the sunlight in between, and its peak radiation is atAt 340nm. UVA-340 for different experiments

Comparative testing is particularly useful in shortwave ultraviolet radiationlineBand can bebestSimulate sunlight.Reference point for UVA-340 irradiance setting:

0.68W/m2 is equivalent to the intensity of noon sunlight in summermaximumValue, get quick results.

9.4.UVB-313fluorescent tube

The wavelength range isBetween 280nm and 315nm. The peak is at 313nm. UVB-313 causes material aging compared to UVA-340 lamps

fasterBut their wavelengths shorter than the cut-off point of sunlight may be impractical for many materialsknotFruit.UVB-313 radiation

Reference point for illuminance setting:0.67W/m2 is a typical light intensity for UVB-313 lamps, which can be tested very quickly and used for quality control or

Testing of materials with good durability.

9.5.UV cycle temperature control

The components of the UV circulation system include a blower, an air heater, an air distribution system, and a blackboard temperature sensor.

aBlower: installed at the air duct of the lower box of the equipment, throughout the entireIn the UV cycle, the blower works continuously.

bAir heater: Installed in the air duct above the blower, it is used to heat the air blown out by the blower when needed.

cAir distribution system: The air blown out by the blower enters the laboratory through the air duct at the center of the water tray, and is then guided by air

Guide to fill the entire laboratory.

9.6Condensation system(Simulated Dew)

The condensation system is essentially the evaporation of water vapor confined within the experimental chamber by the sample of the experimental chamber. During the condensation cycle, water passes through

The heating element in the plate heats up, and water vapor fills the test chamber100%When the humidity reaches equilibrium, water vapor continuously condenses on the sample

Finally, it flows back into the water tray.

9.7Spray system(Simulating rainwater)

The sprinkler system is equipped with a high-power water pump,It consists of 8 nozzles (4 on each side), connecting pipes, control valves, and drainage parts. Installation of nozzle

Between the lamp tubes, in the spray cycle, water is sprayed onto the sample to produce thermal shock or erosion effects.

tenCharacteristics of box structure

10.1. Equipment sample area

The outer shell of the test chamber is equipped with door opening devices at each opening, making it easy to open.

For ease of movement, casters are installed at the bottom of the instrument and fixed in position before useFix the position of the testing machine with a U-shaped ring.

10.2.Inner box material

The material of the inner panel ispremiumSUS304 stainless steelHas good heat resistance, low temperature strength, and mechanical properties; Stamping, bending, etc

Good heat workability, no heat treatment hardening phenomenon (non-magnetic, operating temperature)-196℃~800℃)。

10.3.outsideBox material

The box is made of fully CNC technology, with smooth lines and a beautiful and elegant appearance,Exterior materialFine powder baking paint treatment.

tenonecontrol system

11.1Controller (model)

self-developedprogrammable7-inch true color touch screen controller

11.2. Programway

lOperation mode: program mode, fixed value mode

lSetting method: Chinese menu, touch screen input

lInterface language type: Chinese/English

lProgram editor: editable120 sets of programs, each set of programs can write up to 100 segments and can be set to run in a loop;

lBlackboard temperature monitoring: using standardPt-100 blackboard temperature sensor accurately controls the surface temperature of the sample during the testing process.

lCommunication interface:(RS232/RS485, The longest communication distance is 1.2km [fiber optic can reach 30km]

lAuxiliary functions: fault alarm and cause, processing prompt function, power-off protection function, timing function(Automatic start and automatic)

Stop running)Self diagnostic function.

tentwoUsage conditions (guaranteed by the user for the following conditions)

12.1. Environmental conditions

lPower requirements:220V ± 5%, single-phase threeline50Hz,8A, Require 10A slow melting fuse.

lenvironment5~35 ℃, 0~80% RH, good ventilation, clean indoor environment.

lLand occupation for work: approximately250×360cm

lDrainage: There needs to be a drainage ditch on the floor near the host.

lFor ease of movement, casters are installed at the bottom of the instrument and the position is fixed

lLater useFix the position of the testing machine with a U-shaped ring.

12.2.installation site

lThe ground is flat, well ventilated, and free of flammable, explosive, corrosive gases, and high concentration dust

lThere is no strong electromagnetic radiation source nearby

lNo direct sunlight or direct radiation from other heat sources

lLeave appropriate maintenance space around the equipment

12.3.Maximum current power consumption

Current:(20)A Power:4.5KW Voltage RangeNot exceeding ± 10%

12.4.Shipping must be accompanied by

lOne copy of the product factory certificate of conformity

lOne user manual for the product

lwarranty card

lOne user manual for control instruments