Japan RION KC51 dust particle counter in stock
Japan RION KC51 dust particle counter in stock
KC-51 Handheld Dust Particle CounterIt is produced by Nippon Riken Corporation and weighs 780 grams. It adopts a new streamlined body that allows sample air to freely flow around the body, enabling reliable measurements. The body of the human power design has a large and easy to observe display screen, which is convenient for use
Nippon Riken KC-51 Laser Dust Particle Counter
Nippon Riken KC-51 Dust Particle CounterParticle size range: 0.3, 0.5, and 5.0 μ m. 3.2-inch monochrome LCD panel. There is a USB terminal designed on the instrument panel
KC-52/KC-51Both handheld laser particle counters comply with the ISO 21501-4 standard (JIS B 9921). Can be used in various pollution control applications.
The Nippon Riken KC-51 laser dust particle counter is already in use in the industry:
Environmental control in the preparation and packaging processes of pharmaceuticals, beverages, and food.
Can connect to multi-point monitoring systems.
Pollution control in clean rooms.
Environmental control in semiconductor manufacturing plants.
Pollution control of hospitals and medical equipment.
Technical parameters of Nippon Riken KC-51 laser dust particle counter:
Product Model |
KC-51 Laser Dust Particle Counter |
optical system |
Lateral light scattering method |
light source |
Laser diode (wavelength 780nm, rated output 35mW) |
Laser product grade |
Class 1, compliant with IEC60825-1 |
photodetector |
photodiode |
Sample flow rate |
283L/min |
Particle size range |
User selectable channel count: 3 channels: ≥ 0.30 μ m, ≥ 0.50 μ m, ≥ 5.0 μ m (default setting) 2 channels: ≥ 0.30 μ m, ≥ 0.5 μ m 2 channels: ≥ 0.50 μ m, ≥ 5.0 μ m |
Particle number concentration |
70000 particles/liter (compliant with a loss of 5%), 140000 particles/liter (compliant with a loss of 10%) |
Measurement time/measurement volume |
人工 0.283L(6s),1L(21s),2.83L(1min),10L(3min,32s),28.3L(10min) |
Calculation display mode |
Cumulative |
Set the volume for transformation |
L,28.3L,1000L, none |
Continuous measurement number |
~10 |
Measurement interval |
~10 seconds (set as sleep cycle) |
The number of measurements that can be saved based on the measurement results |
~100 times (based on rotation) |
Saved data format |
Text file in TSV format |
Save data retrieval method |
Download to a connected personal computer (using specialized software) |
monitor |
3.2-inch monochrome LCD panel. |
Operation Method |
button |
use language |
English |
external terminal |
USB Mini B (for connecting to a personal computer or printer) |
Data connection terminal (for connecting to multi-point monitoring system *) |
|
power supply |
AC adapter (input voltage: 100-240V AC, 50/60Hz) Lithium ion battery pack (built-in) |
size |
304(H)×86.6(W)×45.1(D)mm |
weight |
Approximately 780g |
The attachments provided |
AC adapter power cord portable instrument box countdown time (zero) filter instrument handheld strap equal power probe TSV download software |
Optional accessories |
USB thermal printer cable printer table sampling tube (outer diameter 8mm and inner diameter 6mm) holder USB cable (A-Mini B) |