Handheld particle counter 3888/3889 with field Kanomax
Dust particle counterBefore use in clean rooms or other clean environments, please first clean the interior of the product with the zero filter that comes with it.
Install a zero point filter on the air inlet,
Turn on the power, click the start button, and perform continuous measurements for about 15 minutes,
The count value is stable and cannot show any increase or decrease,
This state must be maintained for more than 10 seconds before measurement can be taken.
Remove the zero point filter.
Set necessary items through the general settings menu and measurement
And the measurement conditions can also be determined from the preset.
After returning to the main interface, click the start button to start the measurement
The measurement can be automatically or manually stopped by clicking the stop button to end the measurement.
Through the resume menu, past measurement results can be confirmed

Can simultaneously detect six particle sizes (0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μ m)
-Sampling volume 2.83L/min
-4.3-inch true color touch screen
-Four language switching display (Chinese, English, Japanese, Spanish)
-Large storage capacity and data statistics function
-External sensors can be used to test environmental temperature and relative humidity
-The counting report complies with GB/T 16292-2010, ISO14644-1, ISO 21501-4, and the new version of GMP standards in China
-Built in multiple testing methods (single, repeated, continuous, calculated, ISO, GB)
-Equipped with USB and WIFI communication modes for convenient data transfer and setup
| Name | Kanomax dust particle counter | ||
| Model number | 3888 | 3889 | |
| Test particle size | 0.3 、0.5、 5.0μm | 0.3 、0.5、 1.0、 3.0、 5.0、10.0μm | |
| Light source | laser diode | ||
| Traffic volume | 2.83L/min(0.1cfm) | ||
| sampling time | 6sec~99min59sec | ||
| sampling frequency | 1 to 999 times or continuous testing | ||
| Counting alarm | Within the counting range of 1 to 9999999 | ||
| test method | Single/repeated/continuous/calculated/ISO/GB | ||
| bigMeasurable concentration | 2000000 pieces/CF | ||
| Communication | 本体: USBx2(Host、Device) Base: RS485/Ethernet/WiFi |
||
| data storage | 10000 sets of CSV format | ||
| Language | Chinese, English, Japanese, Spanish | ||
| Power supply | 2 rechargeable lithium batteries (3.7V-2600mAh), AC adapter 100-240V | ||
| Battery working time | 5 hours (single test) | ||
| Overall dimensions | 102(W)×56(D)×215(H)mm | ||
| Weight | About 650g | ||
| runtime environment | 10-40 ℃, 0-85% RH (no condensation) | ||
| Attachment | Testing software, zero count filter, 2 rechargeable lithium batteries, AC adapter, USB cable, air inlet protective cap, quick instruction manual, base, carrying case | ||
| Select items | Temperature and humidity probe, printer, tripod | ||
| Name | Temperature and humidity sensor | ||
| Model number | 0842 | ||
| Range and accuracy | temperature | 0~50℃ ±0.5℃ | |
| humidity | 2.0~98% RH 30~80% RH accuracy ± 3%, accuracy ± 5% for other humidity ranges | ||
Handheld particle counter 3888/3889 with field Kanomax