The electrolytic chlorine gas moisture analyzer is a specialized tool used to determine the moisture content in chlorine gas. In fields such as industrial production, environmental monitoring, and scientific research, it is often necessary to measure the moisture content in chlorine gas. Because the moisture content directly affects the quality of chlorine gas, which in turn affects production and the environment. And it can accurately measure the moisture content in chlorine gas and control it reasonably.
The principle of the electrolytic chlorine gas moisture analyzer is to inject chlorine gas into an anode chamber containing an electrolyte solution, and through the ionization reaction in the electrolyte solution, decompose the moisture in the chlorine gas into hydrogen gas and chlorine gas. The decomposed hydrogen gas forms hydrogen bubbles on the anode, and the number of bubbles represents the moisture content in the chlorine gas. By calculating the number of bubbles, the moisture content in chlorine gas can be accurately determined. The measurement results are intuitive, fast, and accurate.
There are many advantages to using an electrolytic chlorine moisture analyzer. Firstly, it does not require the use of any chemical reagents, thus it does not generate any environmental pollution and does not consume any chemical reagents. Secondly, the use of this measuring instrument is very simple. Simply inject chlorine gas into the anode chamber to quickly obtain the measurement results. Finally, the electrolytic chlorine gas moisture analyzer is very accurate and can determine the moisture content in chlorine gas in a short period of time.
Of course, there are also some issues with the electrolytic chlorine moisture analyzer. For example, it can only measure the moisture content in chlorine gas and cannot measure other components. At the same time, the hydrogen gas generated in the anode can also affect the accuracy of the measurement results, so certain corrections need to be made during use.
In summary, the application of electrolytic chlorine gas moisture analyzer in modern production can improve the quality of chlorine gas and ensure the safety of production and the environment. In the future, with the continuous development of technology, the electrolytic chlorine gas moisture analyzer will also be further improved and developed, bringing more accurate and practical measurement results.
Analysis principle of TMA-210-P chlorine gas moisture analyzer using German CMC electrolysis method:
The P2O5 sensor utilizes the principle of electrolyzing water molecules into hydrogen and oxygen. This sensor consists of a glass cylinder and two parallel electrodes. The electrode material is selected according to the specific application (usually made of platinum or rhodium metal wire), and a thin layer of H3PO4 phosphate film is coated between the two electrodes. The electrolysis current that occurs between the two electrodes decomposes the water in the acid into H2 and O2. The final product of this process is P2O5, which is a highly hygroscopic substance. Therefore, water is absorbed from the sample gas. Through a continuous electrolysis process, a balance is established between the water content of the sample gas and the water content after electrolysis. The electrolysis current is proportional to the water content of the sample gas, and the signal is amplified and displayed. And read out the data. This principle is used to measure Xe、Ar、Kr、He、D2、F2、N2、H2、O2、O3、HBr、PH3、SF6、Freon、C2H2、CO2、CH4、Natural gas, Especially suitable for measuring trace moisture in high-purity acidic gases such as Cl2, HCl, SO2, H2S, and many other gases (except for a few gases that undergo chemical reactions with phosphoric acid).
Technical features of TMA-210-P chlorine gas moisture analyzer using German CMC electrolysis method:
1. Suitable for moisture analysis of corrosive gases and flammable gases
2. Measurement method, no need for recalibration, if necessary, regenerate the sensor electrode film layer
3. Ultra fast response
4. High sensitivity
5. Fully microprocessor controlled
6. Automatic range switching
7. Instrument startup self-test function
8. Complies with NAMUR standards
Main technical parameters of the German CMC electrolytic chlorine moisture analyzer TMA-210-P:
Range: 0~10ppm/100ppm/1000ppm/2500ppm (automatic range)
Accuracy: ± 1% of the full range
Sensitivity: ± 0.1% of the full range
Display: LCD display, backlight switchable
Alarm output: 1 alarm relay
Analog output: 1 channel 0/4~20mA or 0~10V output, users can set it through the menu
Power supply: Built in rechargeable lead-acid battery or external mains power of 80~230VAC/50~60Hz
Power consumption: 8W
Environmental temperature: -10 ℃~+50 ℃
Shell: Protection level IP20
Size:
Width 363mm x Height 115mm x Depth 377mm
Weight: Approximately 6Kg