E-mail
hnfrep@163.com
-
Phone
13611654820
-
Address
Building 3, No. 999 Dongye Road, Sijing Town, Songjiang District, Shanghai
Dalk (Shanghai) Instrument Equipment Co., Ltd
hnfrep@163.com
13611654820
Building 3, No. 999 Dongye Road, Sijing Town, Songjiang District, Shanghai
| characteristic: | The pH dissolved oxygen dual channel controller-M600 can measure parameters such as pH/ORP, dissolved oxygen, temperature, etc., and is used in industries such as pharmaceuticals, chemical engineering, and biological fermentation. |
| Brand: | DALCK |
| model: | M600 |

PH Dissolved Oxygen Dual Channel Controller - M600 Features:
1. It can measure multiple parameters such as pH/ORP, dissolved oxygen, and temperature.
2. Aluminum shell explosion-proof design, IP67 protection level, suitable for various sites.
3. Large screen LCD display with bright backlight.
4. Icon Chinese/English menu, intuitive display, easy and user-friendly operation.
5. Password protection to enhance security protection.
6. Two pH measurements can be selected to meet different on-site needs of users.
7. Three independent multifunctional relays are set up to meet the precise control needs of users.
8. Optional Ethernet communication.

PH Dissolved Oxygen Dual Channel Controller - M600 Technical Parameters:
| PH measurement range | 0.00 to 14.00pH |
| PH resolution and accuracy | 0.01pH and ± 0.02pH |
| ORP measurement range | ±1500mV |
| ORP value resolution and accuracy | 1mV and ± 1mV |
| Temperature measurement range | 0~150℃ |
| Resolution and accuracy of temperature values | 0.1 ℃ and ± 0.2 ℃ |
| temperature compensation | 0-150 ℃ manual/automatic |
| Temperature input | PT100/PT1000/NTC10K/NTC22K/NTC30K |
| DO measurement range | 0~400%/0~40mg/L |
| DO value resolution and accuracy | 0.01mg/L/0.1% and ± 0.2mg/L |
| PH input impedance | ≥1012Ω |
| input power | 22-256VAC/DC |
| Operating Temperature | Minus 10 ℃~60 ℃ |
| signal output | 2-channel 4-20mA isolation protection output |
| relay output | 3-channel multifunctional relay |
| contact method | RS485 communication |
| instrument size | 160 * 160 * 85mm (length x width x height) |
| Installation method | Panel installation |